Using absolute pressure sensors to measure hydrostatic level


27th October 2007

Measuring hydrostatic liquid level is normally carried out using a vented gauge reference pressure sensor. However, venting the inside of a pressure sensor to the outside world is not always possible and when it is, failures can occur due to internal condensation particularly with submersible pressure sensors and loggers.

When measuring for example the water level in a freshwater reservoir, the hydrostatic pressure sensor is measuring both the pressure of water above its position and the atmospheric air pressure acting on the surface of the reservoir.  Normally a vented reference would be used to equalise the offset due to atmospheric pressure so that the true water level is measured.  However if it is necessary to use an absolute hydrostatic pressure sensor this is not possible and the atmospheric pressure has to be measured at the surface and taken away from the reading of the absolute hydrostatic pressure sensor.

In geotechnical applications for example where there are several submersible borehole level loggers or sensors installed within a local area, only one barometric pressure sensor is necessary to compensate all the other level instruments.  This can work out much lower cost especially with deep boreholes where the cost of vented cable and desiccant packs outweighs the cost of one extra barometric pressure logger or sensor.

Where there is a ratio greater than 10 to 1 between the hydrostatic pressure and the barometric pressure no significant additional errors will be introduced over the vented gauge reference method as longs as both pressure instruments are of a similar accuracy grade.  But when the liquid level range falls below 100 metres, the barometric pressure compensation starts to play a bigger part in the error budget eventually increasing the errors effectively by 300% for a 10 meter level range for instance. 

This disproportionate increase in accuracy is caused by having to use a 2 bar absolute range to measure a 10 metre (approx 1 bar gauge pressure) level range plus a barometric range of approximately 1 bar absolute.  So if it is assumed that each sensor has an accuracy of 0.1% full scale the overall error will be 0.3% full scale when compared to a vented hydrostatic level instrument with a level range of 1 bar gauge (10 mWG). 

The atmospheric pressure is effectively being measured twice and as the level range reduces it becomes the dominant component of the total error.  This becomes even worse with level ranges between 1 and 10 meters of water and quickly the economical and practical benefits of using an absolute hydrostatic pressure instrument become over-shadowed by the severe degradation in accuracy.



Pressure Measurement Resource Links

Convert a pressure unit

Convert an altitude in feet or meters to a pressure unit

Convert temperature units to and from degrees celsius, fahrenheit or kelvin

Explain an IP rating for ingress protection

Explanations for pressure measurement terms

Create a pressure sensor specification



Home > News >

Request Info | Ask a Question | Call Request

Site Last Updated:  12/05/08

Tools
Pressure Conversion


Equals

Temperature Conversion


Equals

IP Ratings Explained

I P

First Digit (Solids)
No protection

Second Digit (Liquids)
No protection

Resources

Pressure Measurement Glossary - Find explanations of terminology used for specifying and using pressure instruments

Pressure Unit Conversion - Look up conversion factors or select a conversion table for a particular pressure unit

Pressure Sensor Specification - Select parameters with PSAT to generate a spec for your own use or to submit as an enquiry

Contact

Tel: +44 (0)1780 721569

Fax: +44 (0)1780 721586

Email: Enquiries

Help:  Request Info

         Ask a Question

         Call Request

News
RSS News Feed
RSS

©2008 SensorsONE Ltd, All Rights Reserved.

Powered by interdesigns