There a number of technologies out there for sensing low range pressures such as variable capacitance, variable inductance and strain gauge, but what they all have in common is a reliance on a diaphragm to collect the force and convert it to an electrical signal of some kind. In order to measure small pressure changes the diaphragm has to have a sufficiently low thickness to surface area ratio in order to provide enough flexibility and thus electrical signal to be measured.
Thin diaphragms have lower natural resonant frequencies than thicker ones and therefore lower range pressure sensors are more vulnerable to shock and vibration which can cause large zero shifts in the electrical signal effecting calibration and in extreme cases the diaphragm maybe destroyed by a sudden shock. On the whole though the majority of sensors are robust enough for most applications and it is mainly the effects on accuracy performance that need to be considered.
One acceleration effect which is present in every installation is the acceleration due to gravity. If you consider a thin and large diameter diaphragm orientated in a horizontal plane, it will sag slightly downwards due to the effects of gravity acting on its own weight. If you now flip the diaphragm over, the opposite side will now sag and if you happen to be monitoring the output signal you would have noticed a positive and negative change in the zero pressure output due to what is called the ‘g’ effect. So unless the low pressure sensor happens to be part of an orbiting spacecraft, care should be taken to install a low pressure sensor in the same orientation it was calibrated. If this is not possible, adjustments should be made to the sensor or associated measurement instrumentation to null the zero offset once installed.
External mechanical stress can be picked up by a low pressure sensing diaphragm if it is closely coupled with the surrounding components that make up the rest of the pressure sensor’s construction. For example if the diaphragm is located close to a pressure connector, there maybe some coupling of material stresses when tightening during installation which will manifest as a zero offset in the output signal. Choosing a pressure sensor where the sensing diaphragm is isolated from the pressure connector ports, such as via some internal flexible tubing, will ensure that none of the stresses applied during fitting will cause a calibration shift.
Most pressure sensing devices include thermal compensation circuitry to take out a large portion of the errors caused by thermal expansion of the sensing element components. For low pressure sensors these temperature effects are exaggerated due to the high sensitivity of the diaphragm to any form of induced mechanical stress. Ideally low pressure sensors should be positioned where the temperature is most constant and closest to room temperature. If this is not possible a short length of dead ended pipe should be fitted to provide a ‘stand off’ for the sensor, so that the media inside the pipe cools to ambient temperature before reaching the sensor.
Related Products
Low range pressure instrumentation
Associated Product Articles
Low cost low differential pressure sensor for HVAC applications
Low pressure sensor for use on liquids
Submersible level sensor for shallow water
Site Last Updated: 09/02/12
Glossary - Find explanations for terminology used in specifying and using pressure instruments
Pressure Conversion - Look up conversion factors or select a conversion table for a particular pressure unit
Pressure Measuring Guide - Collection of guidance notes for sourcing, setting up and using pressure measurement equipment
News Articles - Archive of application notes, product releases and technical information
Answers - Search for questions already answered or submit a new question about a pressure measurement topic or product
Pressure Sensor Specification - Select parameters with PSAT to generate a spec for your own use or to submit as an enquiry
Product Finder - Choose multiple product and application types to filter out a suitable product
©2012 SensorsONE Ltd, all rights reserved.