TKDA Digital Auto-Zero Pressure Transmitter

TKDA Digital Auto-Zero Pressure Transmitter

Click Images to Enlarge

Contactless magnetic coupling zero tare & span adjustment pressure transmitter with 4-20mA current loop or amplified voltage output signal in pressure ranges from -1 to +1 bar up to 500 bar gauge

Quote | PDF | Spec | Links 

The TKDA digital auto-zero & span pressure transmitter incorporates an innovative and versatile zero & span trimmer which removes the need to manually adjust potentiometers or purchase expensive proprietary handheld terminals to set the 4 to 20mA current output.

All pressure transmitters drift over time and the main component of drift is contributed by the change in zero offset. The Digital Autozero makes it very easy to perform a zero adjustment as often as necessary to ensure the most accurate pressure measurement.

Digital AutoZero & Span works by simply placing a magnetic pen close to a marked target on the body tube of the pressure transmitter and by following the calibration procedure steps the Pressure transmitter can be calibrated very quickly without any fiddling with pots or connecting handheld computers or PDAs.

The 4-20mA output zero can be coarse trimmed up to +/-10% Full Scale, and fine trimmed to a resolution of +/-0.075% Full Scale.

The TK automation pressure transmitter has a robust Inox AISI430F diaphragm with thick film strain gauges that is ideal for punishing high duty applications such as pneumatic compressors, hydraulic presses, mobile hydraulics, trains, construction vehicles, trucks, system integration, panel builders, & control systems.

 

Technical Summary

Pressure Ranges: +/-1 up to 500 bar gauge and compound pressure

Accuracy: 0.5% FS or improved 0.25% FS

Output Signal: 4 to 20 mA current or amplified voltage

Electrical Connections: Plug and cable options with IP65 to 67 protection rating

Pressure Connections: BSP, NPT and metric thread

Special Features: Auto zero & span setting


Related Products

Pressure calibration equipment

Hand pumps for applying calibration pressures

Auto zero and span pressure transmitter for ranges above 500 bar


Resources

Pressure sensor technical terms explained

Convert pressure into different units


Download Product Information

TKDA data sheet TKDA data sheet

TK data sheet TK data sheet

TKDA operating manual TKDA operating manual


Request a Quote
Pressure Range:
Output Signal:
Electrical Connection:
Length of Cable (metres) if required:
Pressure Connection:
Accuracy:
Quantity:
Your Name:
Your Email:
Comments:

Home > Instrument Type > Pressure Transducers (V) > Mid Range Pressure Transducers with a High Voltage Output

Request Info | Ask a Question

Site Last Updated:  21/08/08

Tools
Pressure Conversion


Equals

Temperature Conversion


Equals

IP Ratings Explained

I P

First Digit (Solids)
No protection

Second Digit (Liquids)
No protection

Resources

Pressure Measurement Glossary - Find explanations of terminology used for specifying and using pressure instruments

Pressure Unit Conversion - Look up conversion factors or select a conversion table for a particular pressure unit

Pressure Sensor Specification - Select parameters with PSAT to generate a spec for your own use or to submit as an enquiry

Contact

Tel: +44 (0)1780 721569

Fax: +44 (0)1780 721586

Email: Enquiries

Help:  Request Info

         Ask a Question

News
RSS News Feed
RSS

©2008 SensorsONE Ltd, All Rights Reserved.